Project Highlight – UCSF Rock Hall Dekker Lab Vac System Replacement

Three large black industrial pumps with exposed piping and wiring installed in a utility room with concrete floors, visible ceiling pipes, and various electrical panels on the walls.

Project: Rock Hall Laboratory Vacuum System Replacement The original specification and basis of design for the (E) Dekker Laboratory Vacuum System: 675 ACFM @ 28”HGVAC = 45 SCFM @ 28”HGVAC Each Vacuum Pump This equates to approx. 250 vacuum inlets.   The typical laboratory design at 25”HGVAC at the vacuum pump inlet, gives approx. 23”HGVAC […]