More Projects
Sutter Roseville “Cart Wash and Ultra Sonic Cleaner” CSPD
December 16, 2021
UC Berkeley Bakar Bioenginuity HUB Laboratory Vacuum System
October 19, 2021
Stanford D&T Mechanical Room Control Air Compressor System Upgrade
September 8, 2021
Stanford School of Medicine Porter Drive Laboratory Vacuum (2) 7.5HP System Installation
Industry: Hospitals and Healthcare, Laboratories
Type: Equipment
Description
Installation of (2) Dekker Titan Single Stage 7.5HP Liquid Ring Vacuum Pump Skids.
Scope of Work
- Provide and Install (1) (N) AES-Dekker Laboratory Vacuum System Model AES-VMX0103MA2-00-DS
- (2) Dekker Titan single stage 7.5HP liquid ring vacuum pump skids.
- Capacity each vacuum pump 15.3 SCFM @ 25” HGVAC.
- System Components Include:
- (2) TEFC Electric motors, 230/460 Volt, 3 Phase, Premium Efficiency.
- (2) Patented DX5 Discharge Separator-Reservoirs with 5 Stage Air/Oil Separation, Including Coalescing Filter.
- (2) Seal Fluid Isolation Valves
- (2) Y-Strainers
- (2) Seal Line Check Valves
- (2) Separator Drain Valves
- (2) Separator Sight Level Gauges
- (2) Temperature RTD for Discharge Temperature Indication and High Temperature Protection.
- (2) Air-Cooled, Heat Exchanger Packages with Main Motor Driven Fan and Fan Guard.
- (2) Inlet Check Valves
- Inlet Manifold Including:
- (2) Inlet Isolation Valves
- (2) Inlet Flex Connectors
- Manifold Piping
- Discharge Manifold Including:
- (2) Individual Pump Drip Legs
- (2) Discharge Check Valves
- (2) Discharge Flex Connectors
- Manifold Piping
- (2) Sets of Hydraulic Hose Seal Fluid Line.
- (2) Inlet Filters
- (2) Inlet Filter Silencers
- (2) Vacuum Relief Valves
- Vacuum Receiver
- Horizontal Grasshopper Vacuum Receiver ASME Coded 120 Gallon.
- Receiver Sight Level Gauge
- Receiver Drain Valve
- Electrical:
- (1) Duplex ControlDEK Control Panel
- Monitors, measures and displays performance
- Identifies maintenance and fault conditions, prevents downtime, and preserves system quality.
- Standard 460V, 3 Phase, 60Hz (NEMA12, CUL Listed)
- Custom PLC with 4.3” Color Touchscreen LCD with integral hour meter
- Across-line Magnetic Vacuum Pump Starter with 3 Phase Overload Protection
- RTD for discharge temperature indication and high temperature protection
- Vacuum On/Off capability with adjustable set points
- Vacuum transducer, for vacuum level indication
- Pressure transducer, for back pressure indication
- 24V Power Supply
- General Fault Dry Contacts
- (1) Duplex ControlDEK Control Panel
Construction-Installation Scope of Work:
- Schedule and Coordinate Shutdown of Existing Laboratory Vacuum System.
- Demo – Remove Existing Airtech Laboratory Vacuum System Model LCS-L-160C-2TAN
- Duplex 7.5HP Oil-Flooded Rotary Vane with Auto Purge
- Electrical Safe off – LOTO to be Completed with Stanford SOM Engineering Support
- FED From MCC1
- Rig Set and Anchor (N) Dekker Oil-Sealed Liquid RingAES-VMX0103MA2-00-DS Laboratory Vacuum System.
-
- Complete Main 480V Power Wiring Terminations (Working with Stanford SOM Engineering)
- Modify Existing Inlet-Suction Piping to Complete Suction POC – Add Temporary Valve
- Start and Test
- Complete System Documentation
Description
Installation of (2) Dekker Titan Single Stage 7.5HP Liquid Ring Vacuum Pump Skids.
Scope of Work
- Provide and Install (1) (N) AES-Dekker Laboratory Vacuum System Model AES-VMX0103MA2-00-DS
- (2) Dekker Titan single stage 7.5HP liquid ring vacuum pump skids.
- Capacity each vacuum pump 15.3 SCFM @ 25” HGVAC.
- System Components Include:
- (2) TEFC Electric motors, 230/460 Volt, 3 Phase, Premium Efficiency.
- (2) Patented DX5 Discharge Separator-Reservoirs with 5 Stage Air/Oil Separation, Including Coalescing Filter.
- (2) Seal Fluid Isolation Valves
- (2) Y-Strainers
- (2) Seal Line Check Valves
- (2) Separator Drain Valves
- (2) Separator Sight Level Gauges
- (2) Temperature RTD for Discharge Temperature Indication and High Temperature Protection.
- (2) Air-Cooled, Heat Exchanger Packages with Main Motor Driven Fan and Fan Guard.
- (2) Inlet Check Valves
- Inlet Manifold Including:
- (2) Inlet Isolation Valves
- (2) Inlet Flex Connectors
- Manifold Piping
- Discharge Manifold Including:
- (2) Individual Pump Drip Legs
- (2) Discharge Check Valves
- (2) Discharge Flex Connectors
- Manifold Piping
- (2) Sets of Hydraulic Hose Seal Fluid Line.
- (2) Inlet Filters
- (2) Inlet Filter Silencers
- (2) Vacuum Relief Valves
- Vacuum Receiver
- Horizontal Grasshopper Vacuum Receiver ASME Coded 120 Gallon.
- Receiver Sight Level Gauge
- Receiver Drain Valve
- Electrical:
- (1) Duplex ControlDEK Control Panel
- Monitors, measures and displays performance
- Identifies maintenance and fault conditions, prevents downtime, and preserves system quality.
- Standard 460V, 3 Phase, 60Hz (NEMA12, CUL Listed)
- Custom PLC with 4.3” Color Touchscreen LCD with integral hour meter
- Across-line Magnetic Vacuum Pump Starter with 3 Phase Overload Protection
- RTD for discharge temperature indication and high temperature protection
- Vacuum On/Off capability with adjustable set points
- Vacuum transducer, for vacuum level indication
- Pressure transducer, for back pressure indication
- 24V Power Supply
- General Fault Dry Contacts
- (1) Duplex ControlDEK Control Panel
Construction-Installation Scope of Work:
- Schedule and Coordinate Shutdown of Existing Laboratory Vacuum System.
- Demo – Remove Existing Airtech Laboratory Vacuum System Model LCS-L-160C-2TAN
- Duplex 7.5HP Oil-Flooded Rotary Vane with Auto Purge
- Electrical Safe off – LOTO to be Completed with Stanford SOM Engineering Support
- FED From MCC1
- Rig Set and Anchor (N) Dekker Oil-Sealed Liquid RingAES-VMX0103MA2-00-DS Laboratory Vacuum System.
-
- Complete Main 480V Power Wiring Terminations (Working with Stanford SOM Engineering)
- Modify Existing Inlet-Suction Piping to Complete Suction POC – Add Temporary Valve
- Start and Test
- Complete System Documentation