Stanford School of Medicine Porter Drive Laboratory Vacuum System Installation

Industry: Hospitals and Healthcare
Type: Equipment

Description

Installation of (2) Dekker Titan Single Stage 15HP Liquid Ring Vacuum Pump Skids w/ Duplex ControlDEK Control Panel.

Scope of Work

  • AES-Dekker Laboratory Vacuum System Model AES-VMXVFD0203KA2-00-ES
    • (2) Dekker Titan single stage 15HP liquid ring vacuum pump skids.
      • 33 SCFM @ 25”HGVAC (Each Vacuum Pump)
    • System Components Include:
      • (2) TEFC Electric motors, 230/460 Volt, 3 Phase, Premium Efficiency.
      • (2) Flexible Couplings
      • (2) Fabricated Steel Baseplates Mounted on Common Base.
      • (2) Patented DX5 Discharge Separator-Reservoirs with 5 Stage Air/Oil Separation, Including Coalescing Filter.
      • (2) Seal Fluid Isolation Valves
      • (2) Y-Strainers
      • (2) Service Liquid Solenoid Valves, 24VDC.
      • (2) Separator Drain Valves
      • (2) Separator Sight Level Gauges
      • (2) Temperature RTD for Discharge Temperature Indication and High Temperature Protection.
      • (2) High/Low Oil Level Switches
      • (2) Air-Cooled, Heat Exchanger Packages with Main Motor Driven Fan and Fan Guard.
      • (2) Inlet Check Valves
      • Inlet Manifold Including:
        • (2) Inlet Isolation Valves
        • (2) Inlet Flex Connectors
        • Manifold Piping
      • Discharge Manifold Including:
        • (2) Individual Pump Drip Legs
        • (2) Discharge Check Valves
        • (2) Discharge Flex Connectors
        • Manifold Piping
      • (2) Sets of Hydraulic Hose Seal Fluid Line.
      • (2) Inlet Filters-Solberg ST-HV Series 3” (Specification and Data Sheet Attached)
      • (2) Vacuum Relief Valves
    • Vacuum Receiver
      • (1) Vertical Vacuum Receiver 200 Gallon.
      • (1) Receiver Sight Level Gauge
      • (1) Receiver Drain Valve
      • (1) Receiver Isolation Valve
    • Electrical:
      • (1) Duplex ControlDEK Control Panel
        • Monitors, measures, and displays performance
        • Identifies maintenance and fault conditions, prevents downtime, and preserves system quality.
        • Standard 460V, 3 Phase, 60Hz (NEMA12, CUL Listed
        • Custom PLC with 4.3” Color Touchscreen LCD with integral hour meter
        • Across-line Magnetic Vacuum Pump Starter with 3 Phase Overload Protection
        • RTD for discharge temperature indication and high temperature protection
        • Vacuum On/Off capability with adjustable set points
        • Vacuum transducer, for vacuum level indication
        • Pressure transducer, for back pressure indication
        • 24V Power Supply
        • (2) VFD’s
        • BacNet and Dry Contacts Included

Description

Installation of (2) Dekker Titan Single Stage 15HP Liquid Ring Vacuum Pump Skids w/ Duplex ControlDEK Control Panel.

Scope of Work

  • AES-Dekker Laboratory Vacuum System Model AES-VMXVFD0203KA2-00-ES
    • (2) Dekker Titan single stage 15HP liquid ring vacuum pump skids.
      • 33 SCFM @ 25”HGVAC (Each Vacuum Pump)
    • System Components Include:
      • (2) TEFC Electric motors, 230/460 Volt, 3 Phase, Premium Efficiency.
      • (2) Flexible Couplings
      • (2) Fabricated Steel Baseplates Mounted on Common Base.
      • (2) Patented DX5 Discharge Separator-Reservoirs with 5 Stage Air/Oil Separation, Including Coalescing Filter.
      • (2) Seal Fluid Isolation Valves
      • (2) Y-Strainers
      • (2) Service Liquid Solenoid Valves, 24VDC.
      • (2) Separator Drain Valves
      • (2) Separator Sight Level Gauges
      • (2) Temperature RTD for Discharge Temperature Indication and High Temperature Protection.
      • (2) High/Low Oil Level Switches
      • (2) Air-Cooled, Heat Exchanger Packages with Main Motor Driven Fan and Fan Guard.
      • (2) Inlet Check Valves
      • Inlet Manifold Including:
        • (2) Inlet Isolation Valves
        • (2) Inlet Flex Connectors
        • Manifold Piping
      • Discharge Manifold Including:
        • (2) Individual Pump Drip Legs
        • (2) Discharge Check Valves
        • (2) Discharge Flex Connectors
        • Manifold Piping
      • (2) Sets of Hydraulic Hose Seal Fluid Line.
      • (2) Inlet Filters-Solberg ST-HV Series 3” (Specification and Data Sheet Attached)
      • (2) Vacuum Relief Valves
    • Vacuum Receiver
      • (1) Vertical Vacuum Receiver 200 Gallon.
      • (1) Receiver Sight Level Gauge
      • (1) Receiver Drain Valve
      • (1) Receiver Isolation Valve
    • Electrical:
      • (1) Duplex ControlDEK Control Panel
        • Monitors, measures, and displays performance
        • Identifies maintenance and fault conditions, prevents downtime, and preserves system quality.
        • Standard 460V, 3 Phase, 60Hz (NEMA12, CUL Listed
        • Custom PLC with 4.3” Color Touchscreen LCD with integral hour meter
        • Across-line Magnetic Vacuum Pump Starter with 3 Phase Overload Protection
        • RTD for discharge temperature indication and high temperature protection
        • Vacuum On/Off capability with adjustable set points
        • Vacuum transducer, for vacuum level indication
        • Pressure transducer, for back pressure indication
        • 24V Power Supply
        • (2) VFD’s
        • BacNet and Dry Contacts Included